Proline Prosonic Flow P 500 HART Use in protective systems
Endress+Hauser 37
Additional
information
Description
The desired simulation value is defined in the Value current
output 1 to n parameter.
Selection
Off
Current simulation is switched off. The device is in normal
measuring mode or another process variable is being simulated.
On
Current simulation is active.
Current output value
Navigation  Diagnostics → Simulation → Current output value (0355)
Prerequisite In the Current output 1 to n simulation parameter, the On option
is selected.
Description Use this function to enter a current value for the simulation. In this
way, users can verify the correct adjustment of the current output
and the correct function of downstream switching units.
User entry 1. Default value: Select 4.0 mA.
2. Default value: Select 20.0 mA.
or
3.59 to 22.5 mA
Additional
information
Dependency
The input range is dependent on the option selected in the Current
span parameter.