Safety
Centrifuge 5430/5430 R
English (EN)
18
3Safety
3.1 Intended use
The Centrifuge 5430/5430 R is designed for separating liquid substance mixtures with different densities,
in particular, for processing and analyzing samples from the human body in in-vitro diagnostic applications
to ensure that the in-vitro diagnostic device can be used according to its intended purpose. This centrifuge
including components is an in-vitro diagnostic device according to Directive 98/79/EC of the European
Parliament and the Council dated October 27, 1998.
Eppendorf centrifuges are exclusively intended for indoor use by trained specialists.
3.2 User profile
This device may only be operated by trained specialist staff. They must have carefully read the operating
manual and be familiar with the function of the device.
3.3 Application limits
3.3.1 Declaration concerning the ATEX directive (94/9/EC)
Due to its design and the environmental conditions inside the device, the Centrifuge 5430/5430 R is not
suitable for use in a potentially explosive atmosphere.
The device only must be used in a safe environment, such as the open environment of a ventilated
laboratory or fume hood. The use of substances which could create a potentially explosive atmosphere is
not permitted. The final decision on the risks associated with the use of these types of substances is the
responsibility of the user.
3.3.2 Maximum service life for accessories
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device to process any substances which may generate an explosive
atmosphere.
WARNING! Risk of injury from chemically or mechanically damaged accessories.
Even minor scratches and cracks can lead to serious internal material damage.
Protect all accessory parts from mechanical damage.
Inspect the accessories for damage before each use. Replace any damaged accessories.
Do not use rotors, rotor lids or buckets with signs of corrosion or mechanical damage (e.g.,
deformations).
Do not use any accessories whose maximum service life has been exceeded.
When inserting the buckets and rotors, ensure that they do not become scratched.