Centrifuge 5430/5430 R
English (EN)
12
2.4 Rotors
The Centrifuge 5430/5430 R can be operated with the following rotors. Before using tubes, observe the
manufacturer's recommended specifications on resistance to centrifugation (max. g-force).
Max. capacity Max. g-force
(rcf)/speed
(rpm) without
adapter
Max.
load per
rotor
bore
(1)
Notes
Acceleration/
deceleration time
(2)
(soft): with soft ramp
Rotor
FA-45-48-11
With aerosol-tight
QuickLock rotor lid
48 tubes 1.5/2.0 mL.
With adapters:
0.2 mL PCR tubes
0.4 mL tubes
0.5 mL tubes
0.6 mL Microtainer
Outer ring:
18,213 × g
Inner ring:
16,048 × g/
12,700 rpm
3.75 g
Aerosol-tight
(3)
QuickLock rotor lid
(aluminum).
≤ 20 s/20 s
≤ 61 s/65 s (soft)
Rotor
F-45-48-11
With polypropylene
rotor lid
48 tubes 1.5/2.0 mL.
With adapters:
0.2 mL PCR tubes
0.4 mL tubes
0.5 mL tubes
0.6 mL Microtainer
Outer ring:
18,213 × g
Inner ring:
16,048 × g/
12,700 rpm
3.75 g
≤ 20 s/20 s
≤ 61 s/65 s (soft)
Rotor
FA-45-30-11
With aerosol-tight
QuickLock rotor lid
30 tubes 1.5/2.0 mL.
With adapters:
0.2 mL PCR tubes
0.4 mL tubes
0.5 mL tubes
0.6 mL Microtainer
20,817 × g/
14,000 rpm
3.75 g
Aerosol-tight
(3)
QuickLock rotor lid
(aluminum).
•PTFE-coated
(particularly resistant
to chemicals),
marked: coated
•Spin columns
available, better with
rotor FA-45-24-11-kit.
≤ 15 s/15 s
≤ 61 s/65 s (soft)
Rotor
F-45-30-11
With polypropylene
rotor lid
30 tubes 1.5/2.0 mL.
With adapters:
0.2 mL PCR tubes
0.4 mL tubes
0.5 mL tubes
0.6 mL Microtainer
20,817 × g/
14,000 rpm
3.75 g
≤ 15 s/15 s
≤ 61 s/65 s (soft)