Product description
Centrifuge 5430/5430 R
English (EN)
13
Rotor
FA-45-24-11-Kit
With aerosol-tight
QuickLock rotor lid
24 spin columns or 1.5/2.0 mL
tubes.
With adapters:
0.2 mL PCR tubes
0.4 mL tubes
0.5 mL tubes
0.6 mL Microtainer
19.090 × g/
13,200 rpm
3.75 g
Aerosol-tight
(3)
QuickLock rotor lid
(aluminum).
Uniquely high edge,
for all commercial
spin columns.
Also observe the note
on centrifugation with
open tube lids (see
Fixed-angle rotors on
p. 33)
≤ 15 s/16 s
≤ 78 s/90 s (soft)
Rotor
FA-45-16-17
With aerosol-tight
QuickLock rotor lid
16 tubes 5.0 mL. 21,191 × g/
14,200 rpm
9.5 g
Aerosol-tight
(3)
QuickLock rotor lid
(aluminum).
≤ 20 s/20 s
≤ 61 s/66 s (soft)
Rotor
S-24-11-AT
With aerosol-tight
QuickLock rotor lid
24 tubes 1.5/2.0 mL. 16,049 × g/
12,700 rpm
3.75 g
Aerosol-tight
(3)
QuickLock rotor lid
(aluminum).
•The rotor must always
be operated with a
rotor lid.
This rotor exclusively is to be
used with 1.5/2.0 mL reaction
vessels. Spin Columns and the
adapters including the
corresponding 0.2 mL, 0.4 mL,
0.5 mL and 0.6 mL vessels
must not be used in this rotor.
≤ 13 s/16 s
≤ 61 s/66 s (soft)
Max. capacity Max. g-force
(rcf)/speed
(rpm) without
adapter
Max.
load per
rotor
bore
(1)
Notes
Acceleration/
deceleration time
(2)
(soft): with soft ramp