EasyManua.ls Logo

Eppendorf Research pro - Technical Data; Pipette Specifications

Default Icon
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
The technical data is valid only when the Research pro is used with Eppendorf pipette tips. The
multi-channel data is valid for eight- and twelve-channel pipettes. The 50 – 1,200 model is
available as an eight-channel version only.
Test conditions in accordance with ISO 8655 for piston-stroke pipettes with an air cushion by
means of a standardized fine balance with a moisture trap.
Number of determinations:
10 pipettings;
degassed distilled water, 20 °C – 25 °C ± 0.5 °C;
maximum speed;
PIP in standard mode of operation (Sec. 5.3);
prewetted pipette tip;
dispensing onto inside of tube.
If the place where the pipette is used is at extremely high altitude, an adjustment must be made
in line with the ambient air pressure
.
Single-channel
Volume range Step size Color Pipette tip Volume Systematic error Random error
µl
µl
code
epT.I.P.S.
µl
%
µl
%
µl
0.5 – 10 0.01
light
gray
20 µl L 1
5
10
± 2.5
± 1.5
± 1.0
± 0.025
± 0.075
± 0.1
± 1.8
± 0.8
± 0.4
± 0.018
± 0.04
± 0.04
5 – 100 0.1 yellow 200 µl 10
50
100
± 2.0
± 1.0
± 0.8
± 0.2
± 0.5
± 0.8
± 1.0
± 0.3
± 0.2
± 0.1
± 0.15
± 0.2
20 – 300 0.5 orange 300 µl 30
150
300
± 2.5
± 1.0
± 0.6
± 0.75
± 1.5
± 1.8
± 0.7
± 0.3
± 0.2
± 0.21
± 0.45
± 0.6
50 – 1,000 1 blue 1,000 µl 100
500
1,000
± 3.0
± 1.0
± 0.6
± 3
± 5
± 6
± 0.6
± 0.2
± 0.2
± 0.6
± 1
± 2
100 – 5,000 10 purple 5 ml 500
2,500
5,000
± 3.0
± 1.2
± 0.6
± 15
± 30
± 30
± 0.6
± 0.25
± 0.15
± 3
± 6.25
± 7.5
Multi-channel
Volume range Step size Color Pipette tip Volume Systematic error Random error
µl
µl
code
epT.I.P.S.
µl
%
µl
%
µl
0.5 – 10 0.01
light
gray
20 µl L 1
5
10
± 5.0
± 3.0
± 2.0
± 0.05
± 0.15
± 0.2
± 3.0
± 1.5
± 0.8
± 0.03
± 0.075
± 0.08
5 – 100 0.1 yellow 200 µl 10
50
100
± 2.0
± 1.0
± 0.8
± 0.2
± 0.5
± 0.8
± 2.0
± 0.8
± 0.25
± 0.2
± 0.4
± 0.25
20 – 300 0.5 orange 300 µl 30
150
300
± 2.5
± 1.0
± 0.6
± 0.75
± 1.5
± 1.8
± 1.0
± 0.5
± 0.25
± 0.3
± 0.75
± 0.75
50 – 1,200 5 green 1,250 µl 120
600
1,200
± 6.0
± 2.7
± 1.2
± 7.2
± 16.2
± 14.4
± 0.9
± 0.4
± 0.3
± 1.08
± 2.4
± 3.6
2 Technical data
Technical data
2
1_E01 bis 04 Seite 8 Donnerstag, 28. Mai 2009 5:17 17

Table of Contents

Related product manuals