About EasyMax Control Solution
Use only with EasyMax test strips.
Write the date you opened the bottle on the bottle label. The control solution is good for three months
from the date the bottle is opened or until the expiration date on the bottle, whichever comes first.
Do not use a control solution that is past the expiration date.
The control solution can stain clothing. If you spill it, wash your clothes with soap and water.
Close the bottle tightly after every use.
Left over control solution should not be added back into the control bottle.
Store the bottle of control solution at room temperature between 36°F - 86°F (2°C- 30°C). Do not freeze.
Note:
Don't drip Control Solution to the entry of the test strip directly. The reagent on test strip might
be drawn into the bottle of Control Solution to cause the degeneration of Control Solution as well
as might contaminate the meter via the test strip.