33
Understanding Your Test Results
The EasyMax Blood Glucose test strips are plasma referenced and calibrated for easier comparison to lab
results. The American Diabetes Association recommends a post-meal glucose level of less than 140 mg/dL
and a pre-meal glucose of less than 100 mg/dL*.
*American Diabetes Association: Diagnosis and Classification of Diabetes Mellitus m(Position Statement).
Diabetes Care 34 (Supp. 1) S66, 2011.
Unusual Test Results
If your test result does not match the way you feel, please follow these steps:
1. Run a control solution test, Chapter 2 "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3 "Testing Your Blood Glucose."
3. If your test results still do not reflect the way you feel, call your doctor immediately.
Note:
1. Extremely high humidity may affect the test results. A relative humidity greater than 90% may
cause inaccurate results.
2. Hematocrit below 30% may cause higher results. Hematocrit above 55% may cause lower
results.