19
Running a Control Solution Test
Please use EasyMax Control Solution with EasyMax V SMBG under Control Solution Mode.
1. Put a test strip into the meter in
the direction of the arrow on
the test strip.
4. Place the meter on a flat
surface, like a table.
5. Remove the control solution
bottle cap and wipe the tip of the
bottle with a tissue.
6. Squeeze the bottle until a tiny
drop forms at the tip of the cap.
2. The meter turns on
automatically and the icons
of and appear.
3. Click (Left Button) to get
into “CtL” mode.