1 Safety Instructions
10 / 62
80110-201_V24169
2022-01
NOTICE
Insufficient acclimatization time, unsuitable temperature during
acclimatization
If the unit was stored or transported below or above a certain tem-
perature, it will take a certain time and temperature to acclimatize.
If you do not observe the rules, the unit can sustain damage and fail.
Acclimatize the unit according to the rules in the Technical Data.
NOTICE
Overheating of the unit due to poor ventilation
If ventilation is poor, the unit can overheat, sustain damage, and fail.
Install the unit in such a way that there is an unobstructed circu-
lation of air around the housing. Installation in confined wall re-
cesses is prohibited.
NOTICE
Penetration of liquid into the unit
The housing is not absolutely watertight. If liquid penetrates, the unit
can sustain damage and fail.
Make sure no liquid can penetrate the unit.
Do not place vessels containing liquids on top of the unit.
Electric shock
WARNING
Defective grounded power outlet, power supply network without
proper grounding, inferior-quality power cord, incorrect line volt-
age, multiple power outlets, extension cords
Risk of electric shock and other injuries to the patient and medical
personnel! Risk of damage to property.
Connect the unit / cart to a properly installed grounded power
outlet.
Only connect the unit to a power supply network with proper
grounding.
Only use the Erbe power cord or an equivalent power cord for this
purpose. The power cord must bear the applicable national test
symbol.
Check the power cord for damage. You must not use a damaged
power cord.
The supply voltage must match the voltage specified on the unit's
rating plate.
Do not use multiple power outlets.
Do not use extension cords.
WARNING
Incorrect power fuse, defective unit
Risk of electric shock to the patient and medical personnel! Risk of
damage to property.

Table of Contents

Other manuals for Erbe APC 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Erbe APC 2 and is the answer not in the manual?

Erbe APC 2 Specifications

General IconGeneral
BrandErbe
ModelAPC 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals