19 / 62
1 • Safety Instructions
80110-201_V24169
2022-01
If you are using accessories from other manufacturers, check
whether the Erbe unit is interfering with other units or being af-
fected by interference itself. You cannot use the unit if there is any
interference.
Damage to the unit and accessories
Alcohol-based spray disinfectant for fast disinfection
With membrane keyboards and paint surfaces there is the risk of
cracks. Propanol and ethanol will erode surfaces.
Do not use these substances.
Alternate use of disinfectant solutions based on different active
ingredients
A color reaction may occur with plastics.
Do not use these substances alternately.
Mix-up of sockets on monopolar socket modules 20140-622,
20140-623
If the sockets are mixed up, the unit will be damaged.
If you use a connecting cable with a monopolar 4 mm dia. connec-
tor, you may only plug the connector into the socket with the blue
ring. The correct socket is marked with an arrow on the illustra-
tion.
Fig. 1-1
Electric load on instrument too high
The instrument can be damaged.
If the damaged area comes into contact with tissue, it can lead to un-
intentional coagulation.
Determine the electrical capacity of the instrument. It is either
printed on the instrument or can be found in the User Manual.