5 Remedying malfunctions
22 / 42
80116-221
10.15
6 A button on the front panel is
pressed when an activation request
occurs.
Do not activate the unit when
making settings on the front
panel.
If the error still occurs, notify
Erbe Service.
7 Restart the unit. If the error
occurs again, notify Erbe Ser
-
vice.
8 Several activation signals are pending
simultaneously.
Actuate only one activation sig-
nal (footswitch or fingerswitch).
Check the correct function of
the fingerswitches and
footswitches.
If the error still occurs, notify
Erbe Service.
9 Activation source (=footswitch) was
not assigned.
The footswitch must
first be assigned to
an output socket
before the footswitch
can be used for acti
-
vation.
Assign the footswitch of the
required output socket.
10 – 18 Restart the unit. If the error
occurs again, notify Erbe Ser
-
vice.
19 After assigning the AUTO START
function, an activation was detected
within a period of 200
ms.
Occurs when an elec-
trically conductive
contact exists
between the elec
-
trodes of the con-
nected instruments
during AUTO START
assignment.
Check the connecting cable to
the forceps or forceps for short-
circuit.
Ensure that there is no electri-
cally conductive contact
between the forcep branches
during AUTO START assignment;
place the forceps on an insu
-
lated base and open the forcep
branches (electrical contact is
interrupted).
If the error still occurs, notify
Erbe Service.
Error
code
Description Additional informa-
tion
Action