31 / 42
6 Maintenance and servicing
80116-221
10.15
From V 1.0.1:
1. Press and hold the blue selection button (COAG) when switching on.
The unit jumps via quick access to the appropriate test program.
2. All relays close. Press the blue selection button (COAG) again. Both
relays (bipolar + monopolar) are closed or opened together. The
switching state of the relays is shown as a symbol in the COAG dis-
play:
COAG display: Bipolar and monopolar relay closed
3. Perform the measurement.
4. Switch the test specimen off.
DC resistance
Testing and measuring equipment
Erbe Art. No. Description
20192-127 Patient cable AE
or
Patient cable AE, international
20192-110
Laboratory measuring cable
20190-045 Electrode handle ICC/ACC
20194-070 Patient cable NE
or
Patient cable NE, international
20194-075
20100-033 Adapter cable NE
Safety tester (with insulation testing >500 V
DC)