5 • Remedying malfunctions
80116-221
10.15
60 + 61 Restart the unit. If the error
occurs again, notify Erbe Ser
-
vice.
62 The user interface settings are not
stored in the EEPROM. Default pro
-
grams are used (and the default set-
tings for Setup and Service).
User programs can-
not be saved.
Attempt to resave user pro-
grams. If not possible, notify
Erbe Service.
63 Error when writing to the external
EEPROM.
Attempt to save the user pro-
gram again. If not possible,
notify Erbe Service.
64 Error when reading the external
EEPROM.
Attempt to read the user pro-
gram again. If not possible,
notify Erbe Service.
70 Restart the unit. If the error
occurs again, notify Erbe Ser
-
vice.
80 + 81 Restart the unit. If the error
occurs again, notify Erbe Ser
-
vice.
94 Microcontroller Flash data is invalid. Invalid software may
have been installed.
Have the mainboard repro-
grammed (by Technical Service
Germany only).
Error
code
Description Additional informa-
tion
Action