Version: 03/02 E - 15 Operation Manual variobike 500
Art-No: 475.403
Operation Manual
variobike 500
Technical Data
Annex A: Technical Data
Model: variobike 500
Operation mode: Continuous operation
Power supply: 230 V, 50 Hz (optional 115 V, 60 Hz)
Fuses: 2 x 0.4 AT at 230V (2 x 0.8 AT at 115V)
Brake principle: Computer-controlled eddy-current brake with torque measurement;
speed-independent as per DIN 13405
Load range: 20 to 600 watt
Speed range: 30 to 130 rpm
Deviation of load : acc. to DIN 13405: max. ± 5% in range 25 to 400 W
however: max. ± 3 W in range 25 to 100 Watt
acc. to manufacturer: max. ± 5% in range 20 to 600 W
however: max. ± 3 W in range 20 to 100 W
Load stages: 5 watt manual, 5 watt or multiples via program
Time interval: 1 to 99 min.
Saddle height adjustment: Continuously variable for statures between 120 and 210 cm
Maximum weight of patient: 120 kg
Displays: LED display
Pulse measurement: 35 – 230 pulse/min
Calibration: Software via keyboard
Calibration protection: Service plug
Electrical inputs: Digital control RS 232: 9 pole, SUB-D
Electrical outputs: Digital: RS 232, 9 pol., SUB-D
Analog: remote start, 6 pole, DIN bushing
Weight: 53 kg
Dimensions: 55 x 100 x 130 cm (W x L x H)
Power consumption: max. 65 VA
Protection class: I
Applied part : Type B according to EN 60601-1
Classification MDD: Class IIa according to 93/42 ECC
Noise level : Less than 70 dbA
Environment operation: Temperature + 10...+ 40 °C
Relative humidity 13...75%, without condensation
Atmospheric pressure 700...1060 hPa