Push the FUNC push-button, the instrument will show the following:
Where :
MEAS shows that the instrument is performing a measurement.
68 shows that the selected standard is the IPTS-68
J shows that the input sensor is a TC type J.
E shows that an external compensation has been selected
25.0 shows the selected compensation value.
XXXXXXX This field displays the measured value in engineering
units.
°C shows the selected engineering unit.
3.3 RTD MEASUREMENTS
NOTE: The instrument will operate in accordance to the selected
temperature standard (IPTS-68 or ITS-90) (see paragraph
"2.3 INSTRUMENT CONFIGURATION").
CONNECTION FOR RTD MEASUREMENTS
HOW TO MEASURE A TEMPERATURE USING AN RTD SENSOR
Push MEAS push-button; the display will show the following:
By pushing s or t push-button, select the desired input type, that
is "RTD".
MEAS 68J E 25.0
XXXXXXX °C
RTD
MEAS
TC RTD mA mV TX
M20-e-3.p65 7/9/01, 2:40 PM23