EasyManua.ls Logo

ero electronic MEMOCAL 2000 - Section 1 General Information; Introduction; Product Specifications

ero electronic MEMOCAL 2000
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4
· Square root extraction and quadratic signal generation.
· Programmable scaling for mA, mV and V measurement and
generation.
· Built in 24 V DC power supply for 2-wires transmitter excitation
and measurement.
· Programmable external cold junction compensation value.
· Peak and valley data hold.
· Direct connection for all input types (no special connectors are
required for TC inputs).
· Up to 50 steps program or smaller programs to total of 50 steps.
· Ramp, soak and step function capability
· 2 dry contact logic inputs for program advance/wait.
1.2 PRODUCT SPECIFICATIONS
1.2.1 GENERAL SPECIFICATIONS
Case: ABS grey case, color similar to RAL 6038.
Case protection: IP 20.
Terminal: 3 external screw FEMALE plugs f 4 mm.
Input protection: all measured and generated ranges are protected
against fault connection to signals up to 30 V AC/DC.
Display: 2 rows by 16 characters alphanumeric back lighted LCD
display
Weight: 600 g. max. (1.4 lb.).
Power supply: 4 batteries AA size :
- Alkaline 1,5 V or
- Ni-Cd 1,2 V or
- Ni-H 1.2 V.
SECTION 1 GENERAL INFORMATION
1.1 INTRODUCTION
The MEMOCAL 2000 is a versatile, portable, hand-held calibrator
developed to solve two different and coexisting customer needs:
field calibration (maintenance) and laboratory calibration (mainte-
nance, quality test, process simulation, training, etc..).
- The field calibration requires small dimensions, no additional
accessories, long battery life, ergonomy, friendly interface, low
temperature drift, high noise immunity and simple programming
capabilities.
- The laboratory requires: interface for data loggers or supervision
and control systems, large range of I/O capabilities, high
accuracy and stability over time.
The features listed below give an idea of the possible applications
and use facility offered by this instrument
· Accuracy 0.015 % (*)
· Temperature drift 0.1 mV/°C
· Operates with standard 1.5 V AA type alkaline batteries or
rechargeable batteries (Ni-Cd or Ni-MH)
· Advanced battery power management for circuit stand-by mode
when not in use.
· 24 hours (average) battery life operation.
· Built in bi-directional RS-232 interface for remote programming
and data recording
· Autoranging during measurement and generation.
· Simulates and measures T/C signals (15 T/C type) and RTD
signals (Pt 100 and Ni 100).
· Generates and measures mA, mV, V and OHM signals.
M20-e-1.p65 7/9/01, 2:39 PM4

Related product manuals