4
· Square root extraction and quadratic signal generation.
· Programmable scaling for mA, mV and V measurement and
generation.
· Built in 24 V DC power supply for 2-wires transmitter excitation
and measurement.
· Programmable external cold junction compensation value.
· Peak and valley data hold.
· Direct connection for all input types (no special connectors are
required for TC inputs).
· Up to 50 steps program or smaller programs to total of 50 steps.
· Ramp, soak and step function capability
· 2 dry contact logic inputs for program advance/wait.
1.2 PRODUCT SPECIFICATIONS
1.2.1 GENERAL SPECIFICATIONS
Case: ABS grey case, color similar to RAL 6038.
Case protection: IP 20.
Terminal: 3 external screw FEMALE plugs f 4 mm.
Input protection: all measured and generated ranges are protected
against fault connection to signals up to 30 V AC/DC.
Display: 2 rows by 16 characters alphanumeric back lighted LCD
display
Weight: 600 g. max. (1.4 lb.).
Power supply: 4 batteries AA size :
- Alkaline 1,5 V or
- Ni-Cd 1,2 V or
- Ni-H 1.2 V.
SECTION 1 GENERAL INFORMATION
1.1 INTRODUCTION
The MEMOCAL 2000 is a versatile, portable, hand-held calibrator
developed to solve two different and coexisting customer needs:
field calibration (maintenance) and laboratory calibration (mainte-
nance, quality test, process simulation, training, etc..).
- The field calibration requires small dimensions, no additional
accessories, long battery life, ergonomy, friendly interface, low
temperature drift, high noise immunity and simple programming
capabilities.
- The laboratory requires: interface for data loggers or supervision
and control systems, large range of I/O capabilities, high
accuracy and stability over time.
The features listed below give an idea of the possible applications
and use facility offered by this instrument
· Accuracy 0.015 % (*)
· Temperature drift 0.1 mV/°C
· Operates with standard 1.5 V AA type alkaline batteries or
rechargeable batteries (Ni-Cd or Ni-MH)
· Advanced battery power management for circuit stand-by mode
when not in use.
· 24 hours (average) battery life operation.
· Built in bi-directional RS-232 interface for remote programming
and data recording
· Autoranging during measurement and generation.
· Simulates and measures T/C signals (15 T/C type) and RTD
signals (Pt 100 and Ni 100).
· Generates and measures mA, mV, V and OHM signals.
M20-e-1.p65 7/9/01, 2:39 PM4