Chapter 3 Commands
■To Monitor Operating Mode Setting
Table 3.9 Operating mode setting monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the operating mode
setting.
<
Example command
>
“1,MODE?”
<
Response format
>
“operation mode code”
1 to 16 MODE?
−
<
Example response
>
“0”
・
The operating mode setting is returned as follows.
0: Constant mode
1: Program mode
■To Monitor Control Mode Setting
Table 3.10 Control mode setting monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the control mode
setting.
<
Example command
>
“1,AMODE?”
<
Response format
>
“control mode code”
1 to 16 AMODE?
−
<
Example response
>
“1”
・
The control mode setting is returned as follows.
1: Unsaturated
2: Saturated
3: Dry/Wet-bulb (For M-type chambers only)
25