Chapter 3 Commands
■To End Operations at Cycle End
Table 3.38 Operation cycle end command
Setting command
Address
Main
command
Optional
parameter
Setting data
Description/Setting data format/Example
command
<
Description
>
Ends operations at the end of the current cycle.
<
Setting data format
>
“SLOWFINISH”
1 to 16 SLOW
FINISH
−
Cycle end
command
<
Example command
>
“1,SLOWFINISH”
■To End Operations Immediately
Table 3.39
Operation
immediate end command
Setting command
Address
Main
command
Optional
parameter
Setting data
Description/Setting data format/Example
command
<
Description
>
Ends operations immediately.
<
Setting data format
>
“QUICKFINISH”
1 to 16 QUICK
FINISH
−
Immediate
end
command
<
Example command
>
“1,QUICKFINISH”
Reference
If this command is sent while pressure inside the chamber is high, steam
inside the chamber is instantly released into the drainage tank, which may
cause a loud sound. Make sure that people in the vicinity do not misinterpret
this to mean trouble.
38