Chapter 3 Commands
3.2 Monitor Commands
Monitor commands have the below format.
Address, main command [, optional parameter]
■To Monitor Chamber Conditions
Table 3.3 Chamber conditions monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the conditions
inside the chamber.
<
Example command
>
“1,MON?”
<
Response format
>
“monitored temperature,monitored humidity,operating
mode,number of alarms occurred”
1 to 16 MON?
−
<
Example response
>
“24.1,100,CONSTANT,0”
・
Operating mode is returned as follows.
Reply Status
OFF:
When control power is OFF
CONSTANT
:
When in the constant mode
RUN
:
When in the program mode
■To Monitor Chamber Model and ROM Version
Table 3.4 Chamber model and ROM version monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the model code of
the chamber and the version of the temperature controller ROM.
<
Example command
>
“1,ROM?”
<
Response format
>
“chamber model code
ROM version”
1 to 16 ROM?
−
<
Example response
>
“EHS-XXX
Ver.
X.XXX”
22