Chapter 3 Commands
To Monitor Program Mode Pattern Total Time
Table 3.17 Program mode pattern total time monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the total amount of
time required to run the current program mode pattern.
Example command
“1,PCHK?”
Response format
“program mode pattern total time”
1 to 16 PCHK?
Example response
“100.0”
* It can take time to calculate the total time required to run patterns that have multiple loops and steps.
If these calculations take 0.1 sec or more, a time-out error occurs. Time-outs should not occur with
most applicable patterns, therefore, if a time-out error occurs with the PCHK? command, check the
pattern from the operating panel and make changes as necessary.
To Monitor Program Mode Total Time Information
Table 3.18 Program mode total time information monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the total time set for
the program mode pattern and the time remaining till pattern end.
Example command
“1,PTIME?”
Response format
“total time (target time),remaining time”
1 to 16 PTIME?
Example response
“100.0,100.0”
*1: If the pattern has yet to start, “0” is returned for the remaining time.
*2: If the program mode pattern total time monitor command is sent prior to issuing this command, the
result of that calculation is returned.