Chapter 3 Commands
3.2 Monitor Commands
Monitor commands have the below format.
Address, main command [, optional parameter]
To Monitor Chamber Conditions
Table 3.3 Chamber conditions monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the conditions
inside the chamber.
Example command
“1,MON?”
Response format
“monitored temperature,monitored humidity,operating
mode,number of alarms occurred”
1 to 16 MON?
Example response
“24.1,100,CONSTANT,0”
Operating mode is returned as follows.

Reply Status

OFF:
When control power is OFF

CONSTANT
:
When in the constant mode

RUN
:
When in the program mode
To Monitor Chamber Model and ROM Version
Table 3.4 Chamber model and ROM version monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the model code of
the chamber and the version of the temperature controller ROM.
Example command
“1,ROM?”
Response format
“chamber model code
ROM version”
1 to 16 ROM?
Example response
“EHS-XXX
Ver.
X.XXX”