Chapter 3 Commands
■To Monitor Humidifying Water Target and Monitored Temperatures
Table 3.5 Humidifying water temperatures monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the target and
monitored humidifying water temperatures.
<
Example command
>
“1,TW?”
<
Response format
>
“target temperature,monitored temperature”
1 to 16 TW?
−
<
Example response
>
“105.0,24.1”
■To Monitor Current Chamber Process
Table 3.6 Current chamber process monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
<
Description
>
This command requests the chamber to return the code for the
current chamber process. Process is determined by chamber
status.
<
Example command
>
“1,SEQN?”
<
Response format
>
“process No.”
1 to 16 SEQN?
−
<
Example response
>
“1”
・
Chamber process is returned as follows.
1: Idling
2: Auto start timer counting down
3: Temperature heat-up
4: Testing in progress
5: Temperature pull-down
6: Exhausting/Draining in progress
7: End
23