Chapter 3 Commands
To End Operations at Cycle End
Table 3.38 Operation cycle end command
Setting command
Address
Main
command
Optional
parameter
Setting data
Description/Setting data format/Example
command
Description
Ends operations at the end of the current cycle.
Setting data format
“SLOWFINISH”
1 to 16 SLOW
FINISH
Cycle end
command
Example command
“1,SLOWFINISH”
To End Operations Immediately
Table 3.39
Operation
immediate end command
Setting command
Address
Main
command
Optional
parameter
Setting data
Description/Setting data format/Example
command
Description
Ends operations immediately.
Setting data format
“QUICKFINISH”
1 to 16 QUICK
FINISH
Immediate
end
command
Example command
“1,QUICKFINISH”
Reference
If this command is sent while pressure inside the chamber is high, steam
inside the chamber is instantly released into the drainage tank, which may
cause a loud sound. Make sure that people in the vicinity do not misinterpret
this to mean trouble.