Chapter 3 Commands
To Monitor Constant Mode Pattern No.
Table 3.11 Constant mode pattern No. monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the current pattern
No. being used in the constant mode.
Example command
“1,CPAT?”
Response format
“pattern No.”
1 to 16 CPAT?
Example response
“1”
To Monitor Constant Mode Parameters
Table 3.12 Constant mode parameters monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the target
temperature, humidity and time set for the constant mode.
Example command
“1,SET?”
Response format
“target temperature,target humidity,target time”
1 to 16 SET?
Example response
“105.0,100,100.0”
To Monitor Constant Mode Time Information
Table 3.13 Constant mode time information monitor command
Monitor command
Address
Main
command
Optional
parameter
Description/Example command/Response format/Example
response
Description
This command requests the chamber to return the target time set
for the constant mode and the time remaining till mode end.
Example command
“1,TIME?”
Response format
“target time,remaining time”
1 to 16 TIME?
Example response
“100.0,50.0”