EasyManua.ls Logo

FBD FBD553D - LCD DISPLAY ERROR MESSAGES

FBD FBD553D
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(Section 14.29 continued from previous page.)
E. Level Control too low. E. Level Control settings may need to
be raised due to excessive CO2 in
the chamber. Raise the Level
Control settings using the
instructions in Section 10.2 of
this manual.
F. "Regulated CO2" transducer F. Insure that the tank transducers
reading inaccurately. are reading the CO
2 regulated
pressure correctly [i.e., filling about
5-7 PSIG (34.5-48.3 KPa) below
“REG CO2”].
G. Secondary regulator G. Insure CO2 regulator has not
pressure creeping up. climbed above preset
28-32 PSIG (193.0-220.8 KPa).
H. CO2 check valve problem. H. Inspect CO2 check valve. Clean
or replace if necessary.
14.30 Max Fill A. Secondary CO2 pressure is A. Check Secondary CO2 for correct
set above 40 psi. setting of 28-32 psi (193.0 - 220.8
KPa).
B. Secondary regulator B. Insure CO2 regulator has not
pressure creeping up. climbed above 32 PSIG. Clean or
repair or replace regulator, if
necessary.
14.31 DefPRESmax A. Tank pressure above 55 psi. A. See 14.29.
14.32 DefProdOut A. Product supply is out during A. Replenish product supply
defrost stage. (i.e., water, CO
2, syrup).
14.33 BTR-LO-ERR A. Product freezing too hard. A. 1. Check brix and correct, if
Beater % below 450. necessary.
2. Check Level Control and
correct, if necessary.
3. Check for restriction of beater
blades.
4. Check for failed beater motor.
B. Bad capacitor on motor. B. Check capacitor and replace as
required.
C. Motor control board failure. C. 1. Check connector from Beater
Motor to lower control board.
2. Change lower control board if
necessary.
14.34 Syrup Out A. Unit receiving no syrup. A. Check to see that syrup container
is not empty and is hooked up.
B. Restriction in lines. B. Assure all lines are free of crimps
or restrictions.
C. Regulator pressures (unit C. Check Syrups pressure readouts
and syrup pumps) too low. for 70-72 psi. Adjust regulators, if
necessary.
D. Syrup Pump failure. D. Check pump and replace,
if necessary.
E. Excessive pressure drops E. Observe “Syrup Press” readouts
in syrup lines. as chamber fills to verify. Increase
line diameter or relocate syrup
source closer to unit. Syrup
pressures may only be raised (by
increasing pump CO2
pressure) if
separate regulators are installed for
unit and syrup pumps. Never
exceed pump manufacturers
recommended maximum CO
2
inlet
pressure (see Section 12.4).
(Section 14.34 continued on next page.)
TROUBLE CAUSE REMEDY
29

Table of Contents

Related product manuals