8 Normalizing
A normalization is used to adjust the measuring instrument to a measuring task.
The
symbol on the display of the instrument indicates the “Normalization” operating mode.
The following items are required for performing a normalization:
Uncoated specimen.
In geometry and substrate material, this specimen should correspond exactly to the parts to be measured.
Note: All readings in the memory will be deleted when performing a normalization.
Normalizing the instrument:
Prerequisite:
Instrument must be turned ON.
1. Select the menu (normalization) with button [] or [].
2. Press button [OK]. The message "Base" (for “substrate (base) material”) appears.