22
11 Setting the specification limits
Example: Lower specification limit (MIN): 20 µm; upper specification limit (MAX): 40 µm
Setting the specification limits:
1. On the display, set the specification limit mode using the arrow key [] or [].
2. Press button [OK]. " - - - " and MIN appear on the display. This means that no values are stored in memory
and that the instrument is ready for entering the MIN value; i.e., it is ready to enter the lower specification limit.
3. Make one measurement on the specimen that is to be tested.
4. Set the desired MIN value using the key [] or []. Example: the lower specification limit 20 µm.
5. Press button [OK] to accept this value into the memory. MAX appears on the display.