20
10 Deleting a calibration / Restoring the baseline of the characteristic
Sometimes, if an instrument does not measure correctly even after a calibration, the calibration can be deleted.
This can occur if the previous calibration has not been carried out correctly.
In such a case, the baseline of the characteristic can be restored to its original factory condition.
Note to users of the DUALSCOPE
MP0R (instrument version with two measurement channels): the base line of
each of the two measurement channels must be restored separately (Fe Base and Al Base required).
Deleting the calibration of the instrument: (Instructions continued on next page)
Prerequisite: Instrument is switched ON.
1. Select the menu option (i.e., calibration) using button [] or [].
2. Press button [OK]. "Base" (i.e., “substrate material”) will appear on the display.
The instrument is now in calibration mode.