26
1) Automatic triggering of the group separator (with fixed block sizes)
Prerequisite: the block size is set > 0.
A group separator is sent automatically after the computation of the block statistics (cf. Chapter 4 “Evaluation of a
test series”, Page 9) and the subsequent automatic deletion of the data of the block in the measurement memory.
2) Manual triggering of the group separator (without creating a block)
Prerequisite: Block size = 0
Obtain any number of measurements.
Select using [] or [].
Press button [OK] to the statistics computation.
Select
using [] or [].
Press button [OK] to delete the measurement memory.
After this delete command, the instrument will send a group separator automatically.