FISCHER DUALSCOPE MP0R USB - Er14 Normalization by Mistake; Er15 Internal Instrument Error; Er17 Not Enough Readings

FISCHER DUALSCOPE MP0R USB
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Normalization performed by mistake on calibration
standard instead of on uncoated specimen.
Perform normalization on uncoated specimen.
Er14 Internal instrument error: Not able to compute
master characteristic. The original master
characteristic is retained.
Repeat master calibration.
If this error occurs repeatedly:
Contact customer service.
Er15 Not able to save the master characteristic.
Probe may be defective.
Contact customer service.
Er17 Not enough readings. Thus, not able to finish a
block (applies to measurements w. fixed block
limits).
While calibration: STD1 has been measured only
once.
Make additional measurements until the first
block can be finished. Then generate a new final
result.
While calibration: STD1 has to be measured for
approx. 5 times.