11 Advanced Measuring Functions FLUXUS F70x
UMFLUXUS_F7V4-6-2EN, 2017-10-01 109
11.10.1 Configuration of the Remote Function
The remote function can be configured for each individual channel.
11.10.1.1 Editing the Remote Function
If the list item Edit is selected in the scroll list of the menu item Remote inputs, the remote function of the selected
channel can be edited.
Resetting the all measured values derived from the flow velocity to 0
Resetting the totalizers
Stopping the totalizers
Select
Special Funct.\SYSTEM settings
\
Remote inputs. Press ENTER.
Select the channel for which a remote function is to be activated. Press ENTER.
In the following scroll list, it can be determined how the remote function of the selected chan-
nel is to be configured or deactivated. Following list items are available:
Edit: editing of the remote function of the selected channel (see section 11.10.1.1)
Copy from: copying the previously configured remote function from a different channel
(see section 11.10.1.2)
Clear all: resetting the previously configured remote functions (see section 11.10.2)
go back: quitting the configuration of the remote function and return to the main menu
Select the list item
Signal Lock. Press ENTER.
The actual measured flow velocity is ignored and the measured value is reset to 0. There-
fore, all values of the physical quantities derived from the flow velocity also yield 0.
When the voltage of the binary input is switched to the passive range, the current mea-
sured values of the flow velocity are again output and used for further calculations.
Select the list item
clear totals. Press ENTER. The totalizers are reset to 0. The total-
izers are deactivated for the duration of the switching signal.
Totalizing restarts at 0 when the voltage at the binary input is switched back to the passive
range.
If the totalizers are reset to 0 with using the remote function, a downward arrow will be dis-
played next to the value of the totalizer during the measurement.
Select the list item
No totalizing. Press ENTER.
The totalizers are stopped for the duration of the switching signal.
Totalizing is continued at the last recorded totalizer value when the voltage at the binary in-
put is switched back to the passive range.
SYSTEM settings
Remote inputs
Remote inputs
for Channel
Remote inputs
A: Edit...
A: Remote-Func.
Signal Lock
A: Remote-Func.
clear totals
A: +0.000 l
88.79 l/h
A: Remote-Func.
No totalizing

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F706**-A2 Series and is the answer not in the manual?

Flexim FLUXUS F706**-A2 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F706**-A2 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals