19 Troubleshooting FLUXUS F70x
UMFLUXUS_F7V4-6-2EN, 2017-10-01 163
19.1 Problems with the Measurement
A measurement is impossible because no signal is received. A question mark is displayed in the lower line on the
right
Check if the entered parameters are correct, especially the outer pipe diameter, the pipe wall thickness and the sound
speed of the medium. (Typical errors: The circumference or the radius was entered instead of the diameter. The inner
pipe diameter was entered instead of the outer pipe diameter.)
Make sure that the recommended transducer distance was adjusted when mounting the transducers.
Make sure that an appropriate measuring point has been selected (see section 19.2).
Try to establish better acoustic contact between the pipe and the transducers (see section 19.3).
Enter a lower value for the number of sound paths. The signal attenuation might be too high due to a high medium vis-
cosity or deposits on the inner pipe wall (see section 19.4).
The measuring signal is received but no measured values can be obtained
An exclamation mark "!" in the lower line on the right indicates that the defined upper limit of the flow velocity is exceed-
ed and, therefore, the measured values are marked as invalid. The limit has to be adapted to the measuring conditions or
checking has to be deactivated (see section 11.4).
If no exclamation mark "!" is displayed, a measurement at the selected measuring point is not possible.
Loss of signal during the measurement
If the pipe had run empty: Was there no measuring signal afterwards? Contact FLEXIM.
Wait briefly until acoustic contact is reestablished. The measurement can be interrupted by a temporarily higher propor-
tion of gas bubbles and solids in the medium.
The measured values substantially differ from the expected values
Wrong measured values are often caused by wrong parameters. Make sure that the entered parameters are correct for
the measuring point.
If the parameters are correct, see section 19.5 for the description of typical situations in which wrong measured values
are obtained.
19.2 Selection of the Measuring Point
Make sure that the recommended min. distance to any disturbance source is observed (see chapter 5, Tab. 5.2).
Avoid measuring points with deposit formation in the pipe.
Avoid measuring points in the vicinity of deformations and defects on the pipe and in the vicinity of welds.
Measure the temperature at the measuring point and make sure that the transducers are suitable for this temperature.
Make sure that the outer pipe diameter is within the measuring range of the transducers.
When measuring on a horizontal pipe, the transducers have to be mounted on the side of the pipes.
A vertical pipe always has to be filled at the measuring point and the medium should flow upward.
No gas bubbles should form (even bubble-free media can form gas bubbles when the medium expands, e.g., upstream
of pumps and downstream of great cross-section enlargements).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F706**-A2 Series and is the answer not in the manual?

Flexim FLUXUS F706**-A2 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F706**-A2 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals