FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA
60.00mA
1) 2)
, 600.0 mA
2) 3)
6.000A
1) 4) 5)
, 10.00A
4) 5) 6)
1) Non-zero residual readings 5d may appear as a zero ampere input when battery indication is 25%
2) A/mA DC accuracy is affected by extreme interior temperature of the meter. For rated accuracies, allow
linear proportional cool down intervals from 6 to 20 minutes after measuring currents from 3 to 10A
continuously via A-input.
3) <400mA continuous; >400mA for <20 minutes ON per >5 minutes OFF
4) 10A continuous up to ambient 35oC; <15 minutes ON per >5 minutes OFF @ 95~122°F (35°C ~ 50°C)
5) >10A to 20A for <30 seconds ON per >5 minutes OFF
6) Unspecified @ <0.5A
Temperature
1) Accuracies assume meter interior has the same temperature as the ambient (isothermal stage) for correct
junction voltage compensation. Allow enough time to reach the isothermal stage for a significant change of
ambient temperature. It can take up to an hour for changes > 9°F (5°C).
2) Type-K thermocouple range & accuracy not included
3) For isothermal stage, allow linear proportional cool down intervals from 9 to 30 minutes after measuring
currents from 3 to 10A continuously via A-input.
Line Frequency