Technical data24
Image presentation modes
Thermal MSX Thermal image with enhanced detail presentation
Picture in Picture Resizable and movable IR area on visual image
Measurement
Object temperature range
• –20°C to +120°C (–4°F to +248°F)
• 0°C to +650°C (+32°F to +1202°F)
Accuracy ±2°C (±3.6°F) or 2%, whichever is greater, at 25°C
(77°F) nominal.
Measurement analysis
Spotmeter
5
Area
5 + 5 areas (boxes or circles) with max./min./aver-
age (in post-acquisition analysis)
Automatic hot/cold detection
Auto hot or cold spotmeter markers within area
Measurement presets No measurements, Center spot, Hot spot, Cold
spot, 3 spots, Hotspot-spot, Hotspot-temperature
Difference temperature Delta temperature between measurement func-
tions or reference temperature
Reference temperature Manually set using difference temperature
Emissivity correction Variable from 0.01 to 1.0 or selected from materials
list
Measurement corrections Emissivity, reflected temperature, relative humidity,
atmospheric temperature, object distance, external
IR window compensation
Colors (palettes) Iron, Rainbow, Rainbow HC, White hot, Black hot,
Arctic, Lava
Alarm
Color Alarm (isotherm) Above/below/interval
Measurement function alarm Audible/visual alarms (above/below) on any se-
lected measurement function
Screening Difference temperature alarm, audible
Humidity alarm 1 humidity alarm, including dew point alarm
Insulation alarm 1 insulation alarm
Set-up
Set-up commands Save options, Programmable button, Reset op-
tions, Set up camera, Wi-Fi, Compass, Bluetooth,
Language, Time & units, Camera information
Service functions
Camera software update Use PC software FLIR Tools
#T559879; r. AO/35410/35410; en-US
97