TROUBLESHOOTING
Symptom Probable Cause Solution
Measurement is
not accurate
1. Turbine operated below
minimum rate
Increase flowrate. See
Specifications.
2. Turbine partially clogged with
dried liquid.
Remove turbine. Clean
carefully. Make sure rotor
spins freely.
3. Turbine bearings partially
clogged with dried liquid.
Remove turbine. Clean
carefully. Make sure rotor
spins freely.
4. Sealant wrapped around
rotor.
Remove turbine. Clear
material from rotor. Make
sure rotor spins freely.
5. Installed too close to fittings.
Install Correctly. See
Installation section
6. Improper connections to
recording device.
Check all electrical
connections. Reference
appropriate installation
instructions.
7. Accuracy needs verification.
Complete normal accuracy
verification procedures.
Repeat periodically.