Questions and Answers

  • L
    Lori WeeksAug 17, 2025
    Why is the display on my Flomec Monitor inaccurate?
    • M
      marshallanitaAug 18, 2025
      An inaccurate display on your Flomec Monitor can stem from several issues. It may be due to an improperly performed field calibration, in which case you should recalibrate or select Factory Calibration, following the Field Calibration Section instructions. Operating the meter below the minimum flow rate can also cause this, so increasing the flow rate might help. Other potential causes include the meter or turbine bearings being clogged with dried liquid (remove and clean carefully, ensuring the rotor spins freely), sealant material wrapped around the meter rotor (remove the rotor and ensure free spinning), or incorrect installation too close to fittings, motors, or an electrically 'noisy' environment (reinstall correctly).
  • T
    travis77Aug 20, 2025
    What to do if the Flomec Monitor display doesn't show count but flowrate is normal?
    • H
      hernandezgrantAug 20, 2025
      If your Flomec Monitor shows a normal flow rate but the display doesn't show a count, it could be due to a few reasons. First, check if the field calibration was performed correctly; if not, recalibrate or select Factory Calibration. Another potential cause is a rotor that's either stuck or damaged. In this case, you should remove the meter and ensure the rotor spins freely. Also, check for any sealant material wrapped around the rotor, and remove it to ensure free movement. If these steps don't resolve the issue, the computer itself might be defective, and you should contact the factory.
  • R
    robertreillyAug 22, 2025
    How to fix a faded or blank display on a Flomec Monitor?
    • J
      james26Aug 23, 2025
      A faded or blank display on your Flomec Monitor usually points to battery issues. Try removing the computer to check and replace the batteries. If replacing the batteries doesn't solve the problem, it's possible that the computer itself is defective, and you may need to contact the factory.
  • T
    tammygravesAug 26, 2025
    What to do if Flomec Monitor has reduced flowrate and the display doesn't show count?
    • T
      Theresa NelsonAug 26, 2025
      If you're experiencing a reduced flow rate and the display on your Flomec Monitor isn't showing a count, it could be because the meter is clogged with dried liquid. Try removing the meter, cleaning it carefully, and making sure the rotor spins freely. Alternatively, the flow rate might be below the minimum required; in that case, increase the flow.
  • K
    Kyle CarrAug 27, 2025
    Why can't I get my Flomec Monitor computer into field calibration?
    • D
      Dr. Brittany RoweAug 27, 2025
      If you can't get the computer into field calibration mode on your Flomec Monitor, it might be due to using the wrong button sequence. Make sure to follow the instructions in the Field Calibration Section. Alternatively, the computer circuit board or a button might be defective. In either of these cases, you would need to replace the computer and contact the factory.
  • B
    Brandon BrooksSep 2, 2025
    What to do if there is no flow on screen when fluid is moving in pipe on Flomec Measuring Instruments?
    • K
      Katherine CalhounSep 2, 2025
      If there's no flow showing on the screen of your Flomec Measuring Instruments when fluid is moving in the pipe, it might be due to a few reasons. You may need to field calibrate again or select Factory Calibration if the field calibration wasn't performed correctly. The rotor could be stuck or damaged, or sealant material might be wrapped around it; in either case, remove the meter and ensure the rotor spins freely. Another potential cause is a defective computer, which may require further assistance.
  • C
    castanedajamieSep 4, 2025
    Why is there excessive flow on screen when fluid is not moving in pipe on Flomec Measuring Instruments?
    • R
      roberta54Sep 4, 2025
      If you're experiencing excessive flow readings on your Flomec Measuring Instruments when fluid isn't moving in the pipe, it's likely due to Electro-Magnetic Interference (EMI). This can happen if the meter is too close to a motor or other electrically noisy equipment. To resolve this, move the installation point of the meter away from EMI sources.
  • J
    Jody PaulSep 6, 2025
    Why is the display faded or blank on my Flomec Measuring Instruments?
    • E
      Emily BullockSep 7, 2025
      If the display on your Flomec Measuring Instruments is faded or blank, the batteries might be weak, dead, or not connected. Remove the display and check/replace the batteries. Alternatively, the ambient temperature might be too cold; place a bare hand or something warmer on the display for temporary readability.
  • J
    Jared FrostSep 9, 2025
    Why is the button not working on my Flomec Measuring Instruments?
    • P
      penny22Sep 9, 2025
      If the button on your Flomec Measuring Instruments isn't working, ensure that your finger covers the entire button when pressing it. Also, the button may only operate upon release, so release the button to see the display change.
  • J
    John BanksSep 12, 2025
    What does “BA0.000” mean on my Flomec Measuring Instruments?
    • T
      Thomas WelchSep 12, 2025
      If your Flomec Measuring Instruments display shows “BA0.000”, it indicates a configuration issue. You can try to reset Batch and Accumulative Totals, or perform a Field Calibration to achieve the desired performance.

Summary

Getting Started

Product Usage and Power Requirements

Details the intended application, power source, and battery requirements for the computer electronics.

General Safety Instructions

Operator Responsibilities and Key Warnings

Outlines operator responsibilities and critical warnings for safe operation and handling.

Specific Hazard Alerts

Details warnings for ignition risk, static discharge, battery hazards, and flammable liquids.

Product Specifications

Mechanical, Electrical, and Temperature Details

Covers physical dimensions, operating temperatures, electrical parameters, and temperature limits.

Power Supply, Features, and Options

Details power supply, battery life, standard features, and optional daughter board modules.

Installation Procedure

Computer Electronics Installation Steps

Provides step-by-step instructions for installing the computer electronics onto a turbine housing.

Operation Overview

Display, Activation, and Normal Operation

Explains the computer display, activation, normal operation, and button functions.

Configuration and Calibration

User Configuration Mode Navigation

Provides general notes for user configuration menus and how to enter the mode.

Calibration Methods and Settings

Describes calibration methods, pipe diameter input, and daughter board module configurations.

Troubleshooting Common Issues

Operational Problems and Solutions

Addresses problems like inaccurate readings, button malfunctions, display errors, and flow anomalies.

Maintenance and Support

Battery Management and Replacement

Provides guidance on battery care, approved types, and the replacement process.

Parts, Service, and Warranty Information

Lists replacement parts, contact info for service, and details the limited warranty.

Flomec Q9 Specifications

General IconGeneral
BrandFlomec
ModelQ9
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals