OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode) (continued)
Field Calibration Method Entry
At the beginning of the calibration method entry menu, the software will allow the user to start
the calibration process or to exit back to normal operation. (See Figures 30-1 & 30-2)
The bottom row of characters will be in focus to indicate that the user can select between
either “Start” or “Exit” by advancing focus. The Rate button changes between START &
EXIT while the Total button makes the actual selection.
NOTE: The Total button must be pressed twice (to verify the end user is "certain" of
selection).
If “Start” is in focus when advancing to the next
menu, the software will automatically advance to
the calibration menu for the calibration method
the user previously selected.
Figure 30-1
If “Exit” is in focus when advancing to the next
menu, the software will display a “save” screen
(see Figure 13) to enter “yes or no” to save your
entries.
Figure 30-2
Dispense Display Method
Dispense/Display is a field calibration method by which the user will dispense a known
volume of fluid and the software will keep track of the pulse count during the dispense
operation. The user will then enter the known volume into the display and the software will
calculate a K-factor for the volume dispensed. (See Figure 31-1)
The software will allow the user to dispense a
known volume in order to create a custom K-
factor. This will not over write the existing factory
calibration table.
Figure 31-1
The first screen in this calibration method will
show “run” indicating that the computer is waiting
for flow to start. (See Figure 31-2)
NOTE: Pressing any single button while on this
screen will not have any effect. Pressing and
holding the Total button first and immediately
pressing and holding the Rate button, holding
both buttons down for 3 seconds, will allow the
user to exit the calibration mode.
Figure 31-2