EasyManua.ls Logo

Fluke Biomedical ProSim 2 - Introduction; Safety Information

Default Icon
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
Introduction
 Warning
To prevent possible electrical shock, fire, or
personal injury, read all safety information
before you use the Product.
The ProSim™ 2 and ProSim™ 3 Vital Signs Simulators
(the Product) are electronic signal sources used to
measure the performance of patient monitors. The Product
simulates:
ECG (with and without arrhythmias)
Respiration
Blood pressure
Temperature
Cardiac output (ProSim 3 only)
Fetal/Maternal ECG and IUP (ProSim 3 only)
The ProSim™ 3 is shown in all illustrations.
Safety Information
A Warning identifies conditions and procedures that are
dangerous to the user. A Caution identifies conditions and
procedures that can cause damage to the Product or the
equipment under test.
Table 1 is a list of symbols used on the Product and in this
manual.
 Warning
To prevent possible electrical shock, fire, or
personal injury:
Use the Product only as specified, or the
protection supplied by the Product can be
compromised.
Do not connect the Product to a patient or
equipment connected to a patient. The
Product is intended for equipment analysis
only. Do not use the Product for
diagnostics, treatment, or other capacity
where the Product touches a patient.
Remove the batteries if the Product is not
used for an extended period of time, or if
stored in temperatures above 50 °C. If the
batteries are not removed, battery leakage
can damage the Product.
Replace the batteries when the low battery
indicator shows to prevent incorrect
measurements.