Vital Signs Simulator
Detailed Specifications
45
Ventricular Arrhythmia ............................................. PVCs (6, 12, or 24 per minute), frequent multifocal PVCs, bigeminy, trigeminy, multiple PVCs (one-time
run of 2, 5, or 11 PVCs), ventricular tachycardia, ventricular fibrillation (coarse or fine), and asystole
Conduction Defect .................................................... First-, second-, or third-degree AV block and right- or left-bundle-branch block
Power-On Default...................................................... None (off)
ECG-Performance-Tests
Amplitude .................................................................. 0.05 mV to 0.45 mV (0.05 mV steps)
0.5 mV to 5.5 mV (0.5 mV steps)
Pulse Wave ................................................................ 30 BPM, 60 BPM, with 60 ms pulse width
Square Wave ............................................................. 2.0, 0.125 Hz
Triangle Wave ........................................................... 2.0, 2.5 Hz
Sine Wave .................................................................. 0.5, 5, 10, 40, 50, 60, 100 Hz
R-wave-Detection Waveform ................................... Haver-Triangle
R-wave Rate .............................................................. 30, 60, 80, 120, 200, and 250 BPM
R-wave Width ............................................................ 20 ms to 200 ms (10 ms steps)
Additional Steps: 8, 10, and 12 ms
Rate Accuracy ........................................................... ±1 %
Amplitude Accuracy ................................................. ±2 %, Lead II (Exception: ±5 % for R waves 20 ms)
Power-On Default ..................................................... None (off)
Respiration
Rate ............................................................................ 0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM
Impedance Variations (Δ Ω) ..................................... 0.2, 0.5, 1, or 3 Ω peak-to-peak variation of lead impedance
Delta Accuracy .......................................................... ±10 %
Baseline ..................................................................... 500, 1000, 1500, 2000 Ω, Leads I, II, III
Accuracy Baseline .................................................... ±5 %
Respiration Lead ....................................................... LA or LL
Apnea Selection ........................................................ OFF, 12, 22, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
Power-On Default...................................................... 20 BrPM, delta 1.0 Ω, 1000-Ω baseline
Blood Pressure
Input/output Impedance ........................................... 300 Ω ±10 %
Exciter Input Range .................................................. 2.0 V to 16.0 V rms