Vital Signs Simulator
Detailed Specifications
47
Temperature
Temperature .............................................................. 0 °C (32 °F), 24 °C (75.2 °F), 37 °C (98.6 °F), and 40 °C (104 °F)
Accuracy .................................................................... ±0.1 °C
Compatibility ............................................................. Yellow Springs, Inc. (YSI) Series 400 and 700
Output Connector ..................................................... DIN 4-pin
Power-On Default...................................................... 0 °C (42 °F)
Cardiac Output (ProSim 3 Only)
Catheter Type ............................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient ............................................. 0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood Temperature ................................................... 37 °C (98.6 °F) ±2 %
Injectate Volume ....................................................... 10 cc
Injectate Temperature .............................................. 0 °C or 24 °C ±2 % value
Cardiac Output .......................................................... 2.5, 5, 10 liters per minute ±5 %
Faulty-Injectate Curve .............................................. Waveform for simulation available
Left-to-Right-Shunt Curve ........................................ Waveform for simulation available
Calibrated Pulse........................................................ 1.5 ° for 1 second (37 ° to 35.5 °)
Output Connector ..................................................... DIN 7-Pin
Power-On Default ..................................................... 2.5 liters per minute, 0 °C injectate
Fetal / Maternal-ECG (ProSim 3 Only)
Fetal Heart Rate (Fixed)............................................ 60, 90, 120, 140, 150, 210, and 240 BPM
Fetal Heart Rate (IUP): .............................................. 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms ........................... Early deceleration, late deceleration, and uniform acceleration
Wave Duration ........................................................... 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0 mmHg
IUP Period .................................................................. 2, 3, or 5 minutes, and manual
Power-On Default...................................................... FHR 120 BPM, early deceleration, manual