Vital Signs Simulator
Remote Operation
31
Table 12. Pacemaker Waveform Commands (cont.)
Action Command
Amplitude
[1]
1 mV PA1
2 mV PA2
5 mV PA5
10 mV PA10
Width
[1]
0.1 mV PA0.1
0.5 mV PA0.5
1.0 mV PA1.0
1.5 mV PA1.5
2 mV PA2.0
1. Set the Pacemaker waveform before you set amplitude and width.
Arrhythmia Functions
Table 13 is a list of the commands to simulate arrhythmias.
These waveforms are grouped by supraventricular
arrhythmia, premature arrhythmia, ventricular arrhythmia,
and conduction defect.
Table 13. Arrhythmia Function Commands
Action Command
Supraventricular
Atrial fibrillation, coarse AF1
Atrial fibrillation, fine AF2
Atrial flutter AFL
Sinus arrhythmia SINA
Missed beat MB80
Atrial tachycardia ATC
Paroxysmal atrial tachycardia PAT
Nodal rhythm NOD
Supraventricular tachycardia SVT
Premature
Premature atrial contraction PAC
Premature nodal contraction PNC
Premature vent contraction left
(PVC1), standard
PVC1S