Accuracy Specifications 
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, 
relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + 
[Number of Least Significant Digits]). 
Function Range Resolution 
Accuracy 
106 107 
AC Volts  
(40 Hz to 500 Hz) 
[1]
 
 
6.000 V 
60.00 V 
600.0 V 
0.001 V 
0.01 V 
0.1 V 
1.0 % + 3  1.0 % + 3 
DC Volts 
 
6.000 V 
60.00 V 
600.0 V 
0.001 V 
0.01 V 
0.1 V 
0.5 % + 3  0.5 % + 3 
AC Millivolts 
 
600.0 mV  0.1 mV  3.0 % + 3  3.0 % + 3 
Diode Test
 [2]
 
 
2.000 V  0.001 V  N/A  10 % 
[1]  All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified. 
[2]  Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA. 
 