EasyManua.ls Logo

Fluke 1520 - Using Lock Function for Insulation

Fluke 1520
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1520
Calibration Manual
12
Maintaining the Meter
This section provides basic maintenance information, including fuse and battery
replacement instructions.
Caution
Do not attempt to repair or service your Meter unless you are
qualified to do so and have the relevant calibration,
performance test, and service information.
Cleaning
Clean only with soap and water. Remove any residue afterwards.
Periodically wipe the case with a damp cloth and mild detergent.
Do not use abrasives or solvents.
Replacing and Disposing of the Batteries
W Warning
To avoid electric shock, disconnect the test leads from the
inputs before opening the Meter for battery replacement.
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the batteries as soon as the
battery empty indicator G
appears.
Note
This Meter contains alkaline batteries. Do not dispose of these batteries
with other solid waste. Used batteries should be disposed of by a qualified
recycler or hazardous materials handler. Contact your authorized Fluke
Service center for recycling information.
The Meter uses four alkaline C cell batteries (supplied). To replace the batteries, do the
following (see Figure 2):
1. Turn the rotary switch to the OFF position.
2. Disconnect test leads from any power source.
3. Remove the holster.
4. Place the Meter face down on a nonabrasive surface and loosen the two screws with a
flat-blade screwdriver.
5. Lift the battery access lid away from the Meter.
6. Replace the C cells as shown in Figure 2. Observe the battery polarity shown in the
battery compartment.
7. Secure the battery access lid back in position with the two screws.

Other manuals for Fluke 1520

Related product manuals