Sets the units to C.
Sets the resolution to AUTO.
Stops auto logging.
Clears the statistical functions.
8.4.8.4 SYSTem:BOOT:VERSion?
Returns the boot ROM version.
SYST:BOOT:VERS?
0.50
8.4.8.5 SYSTem:CODE:VERSion?
Returns the firmware code version.
SYST:CODE:VERS?
1.11
8.4.8.6 SYSTem:ERRor?
Returns a system error message if any are present in the system error queue.
SYST:ERR?
0,“No error”
Each error condition produces only one error message at a time. The error
queue may contain up to ten messages. The messages are reported in the order
they occured. Reading a message from the queue removes the message so the
next message can be read. If more than ten errors occur before being read, the
tenth error in the queue is -350,“Queue overflow” and all but the first nine er
-
rors are discarded. All errors cause the Error Bit (ERR, bit 2) for the Status
BYTE Register to be set. Refer to Troubleshooting section for the error
messages.
8.4.8.7 SYSTem:SNUMber?
Returns the instrument serial number.
SYST:SNUM?
A09002
If the serial number has not been set, a 0 is returned. The serial number is set at
the factory and cannot be changed!
8.4.8.8 SYSTem:VERSion?
Returns the SCPI version number.
SYST:VERS?
1529 Chub-E4 Thermometer Readout
User’s Guide
110

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1529 and is the answer not in the manual?

Fluke 1529 Specifications

General IconGeneral
Channels2
Data LoggingYes
InterfaceRS-232
Resolution0.001°C
Sensor TypePRT (Platinum Resistance Thermometer)
DisplayLCD

Related product manuals