STAT:MEAS:ENAB 15
8.4.12.13 STATus:OPERation?
Reads and clears the Operation Status Register.
STAT:OPER?
16
The assignment of the bits in this register is as follows:
bits 0-3: not used
bit 4: measurement complete
bits 5-15: not used
8.4.12.14 STATus:OPERation:CONDition?
Returns the Operation Status Condition Register.
STAT:OPER:COND?
16
The assignment of the bits in this register is as follows:
bits 0-3: not used
bit 4: measurement in process
bits 5-15: not used
8.4.12.15 STATus:OPERation:ENABle?[MIN|MAX|DEF]
Returns the Operation Status Enable Register.
STAT:OPER:ENAB?
255
8.4.12.16 STATus:OPERation:ENABle <num>|MIN|MAX|DEF
Sets the Operation Status Enable Register.
STAT:OPER:ENAB 16
8.4.12.17 STATus:QUEStionable?
Reads and clears the Questionable Status Register.
STAT:QUES?
16
The assignment of the bits in this register is as follows:
bits 0-3: not used
1529 Chub-E4 Thermometer Readout
User’s Guide