EasyManua.ls Logo

Fluke 1529 - Calibration; PRT and Thermistor Calibration

Fluke 1529
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
bit 4: measurement invalid
bits 5-15: not used
8.4.12.18 STATus:QUEStionable:CONDition?
Returns the Questionable Status Condition Register.
STAT:QUES:COND?
16
The assignment of the bits in this register is as follows:
bits 0-3: not used
bit 4: measurement invalid
bits 5-15: not used
8.4.12.19 STATus:QUEStionable:ENABle? [MIN|MAX|DEF]
Returns the Questionable Status Enable Register.
STAT:QUES:ENAB?
255
8.4.12.20 STATus:QUEStionable:ENABle <num>|MIN|MAX|DEF
Sets the Questionable Status Enable Register.
STAT:QUES:ENAB 10
121
8 Digital Communications Interface
Commands

Table of Contents

Related product manuals