The ECHO parameter allows the user to set the echo (duplex) mode. Selecting
ON enables echoing so that all characters received on the serial port are echoed
back over the serial port. This mode is useful when using terminal emulation
software on a computer to communicate with the instrument. Selecting OFF
disables echoing. This mode is commonly used when using control software on
the computer.
Press the LR buttons to select the parameter setting. Use the UD buttons
to scroll between the parameters. Press the Enter buttontosavethenewset
-
ting. Press and hold the Exit button to cancel to the main display or press the
EXIT button to cancel and to move to the next parameter.
7.5.1.2 GPIB (Optional)
When the GPIB option is selected, the parameters for the IEEE-488 port are
displayed on the bottom portion of the display. These parameters are: GPIB,
ADDR, and TERM.
The GPIB parameter allows the user to enable or disable the IEEE-488 port.
The ADDR parameter allows the user to specify the address of this instrument
on the IEEE-488 bus. The default address is 22 and the range is 1 to 30.
The TERM parameter allows the user to select the termination character trans-
mitted after each transmission. The default termination character is linefeed
(LF).
7.5.2 Date Time
The DATE TIME function is used for setting the time-of-day clock, time for-
mat, date, and date format. When this function is selected, the bottom portion
1529 Chub-E4 Thermometer Readout
User’s Guide
70