EasyManua.ls Logo

Fluke 1750 - Installing the Power Recorder at a Facility; Work Flow

Fluke 1750
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Power Recorder
Verification Tests
11
Table 6. Rogowski Current Accuracy Check
Applied
Voltage
Expected
Current
Expected
Reading
A Reading B Reading C Reading N Reading G Reading Limits
0.649 V 6.49 μA 20 A ± 0.20 A
1.6225 V 6.225 μA 50 A ± 0.35 A
3.245 V 32.45 μA 100 A ± 0.60 A
CT Current Verification
1. Connect the calibrator NORMAL HI to Voltage A, B, and C on the Recorder.
2. Connect the calibrator NORMAL LO to Voltage N and J on the Recorder.
3. Connect the calibrator AUX HI and LO to current input A using the current test
cable. See Figure 6. The current test calibration cable connects the voltage to all
current sense inputs and indicates that a 100 A current transformer is attached.
4. Set the calibrator output to 120.0 V @ 50 or 60 Hz on the NORMAL output and the
values from Table 7 on the AUX output.
5. Press the WAVE MENUS softkey and ensure the phase angle is 0.0 °.
6. Press the HARMONIC MENU softkey; ensure the HARMONIC selection is set to
1 and the FUNDMTL selection is set to aux.
7. Press OPR.
8. Set the calibrator to voltages from Table 7 and list the Recorder readings for that
channel.
Table 7. CT Current Accuracy Check
Applied Voltage
Expected
Reading
A Reading B Reading C Reading N Reading G Reading Limits
0.4 V 20 A ± 0.20 A
1.0 V 50 A ± 0.35 A
2.0 V 100 A ± 0.60 A
9. Verify that the current LED is on.
10. Move cable to next channel and repeat steps 8 and 9 for channels B, C, N, and J.
11. Set the calibrator to STBY. Verify that the current LEDs turn off.

Other manuals for Fluke 1750

Related product manuals