EasyManuals Logo
Home>Fluke>Data Loggers>1750

Fluke 1750 Calibration Manual

Fluke 1750
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
1750
Calibration Manual
10
Neutral Voltage Verification
1. Connect NORMAL LO to A, B, C, J, and ..
2. Connect the calibrator NORMAL HI to N on the Recorder.
3. Apply the voltages from Table 5, listing the Recorder readings for the appropriate
model.
Table 5. Neutral (AC) Voltage Accuracy Check
Applied NG Reading Limits
60 V @ Local Hz ± 1.12 V
100 V @ Local Hz ± 1.20 V
4. Verify that the N LED is on.
5. Set the calibrator to STBY, verify that the LED for N turns off.
6. Disconnect the test leads from the Recorder.
Rogowski Current Verification
1. Verify the resistor in the Rogowski calibration cable is within 0.1 % of 100 kΩ.
2. Connect the calibrator NORMAL HI to Voltage A, B, and C on the Recorder.
3. Connect the calibrator NORMAL LO to N and Voltage J on the Recorder.
4. Connect the calibrator AUX HI and LO to current input A using the Rogowski
calibration cable. See Figure 3.
5. Set the calibrator output to 120.0 V @ 55 Hz on the NORMAL output and the values
from Table 6 the AUX output.
6. Press the WAVE MENUS softkey and ensure the phase angle is -90 °.
7. Press the HARMONIC MENU softkey; ensure the HARMONIC selection is set to
1 and the FUNDMTL selection is set to aux.
8. Press OPR.
9. Set the calibrator to voltages from Table 6 and list the Recorder readings for channel
A.
10. Move cable to next channel and repeat step 9 for channels B, C,
N, and J.

Other manuals for Fluke 1750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1750 and is the answer not in the manual?

Fluke 1750 Specifications

General IconGeneral
BrandFluke
Model1750
CategoryData Loggers
LanguageEnglish

Related product manuals