1750
Calibration Manual
8
Verification Tests
XWWarning
To avoid electrical shock, personal injury, or fire:
Do not perform the calibration procedures or calibration
verification tests described in this manual unless you are
qualified to do so.
Repairs or servicing should be performed only by qualified
personnel.
The following tests are used to verify the functions of the Recorder. If the Recorder fails
any of the verification tests, calibration or repair may be needed. For service, see
Contacting Fluke.
Note
Power Analyze can be set to display the channels as L1, L2, and L3 instead
of A, B, and C. The channels on the Recorder could also be labeled A, B,
and C, or L1, L2, and L3, depending on which decal was applied to the
Recorder front panel. For this manual A, B, and C are used.
AC Voltage Accuracy Verification
1. Connect an Ethernet cable to the Recorder.
2. Apply power to the Recorder. During power up, indicators near all jacks should flash.
3. Launch Power Analyze software on a personal computer (PC).
4. Select the Recorder to be tested in the drop down box next to the Scope button.
5. Select 1750 Live, then Scope.
6. Check all the voltage display check boxes to the right of the scope display. Make sure
the correct items are checked.